Confirm the curve and wiring

Pt100 and Pt1000 have nominal resistances of 100 Ω and 1,000 Ω at 0 °C. Confirm the sensor marking, service range, tolerance class and input type. In two-wire measurement the resistance of both leads adds to the reading. A typical three-wire circuit’s compensation depends on lead matching and the module design. Four-wire measurement separates excitation and voltage sensing, reducing lead resistance error.

Convert resistance using the nominal curve

At t ≥ 0 °C the Callendar–Van Dusen model uses R = R₀(1 + At + Bt²). Below 0 °C the brackets include C(t − 100)t³. The calculator lists the nominal coefficients. A Pt100 measures about 60.26 Ω at −100 °C, 100 Ω at 0 °C and 138.51 Ω at +100 °C on the nominal curve. The corresponding Pt1000 values are ten times larger on the same normalized curve.

Verify the measurement chain separately

Check several points in the operating range using a resistance simulator suitable for the input. Compare the readings against the input module specifications, reference uncertainty and sensor tolerance. Also consider the probe’s operating range, lead resistance and self-heating. A conversion from the nominal curve does not constitute calibration.